Use PT Series Part Number Crossover Tool to convert from 100, 200, 300 and 615 Series transmitters to new PT Series transmitters.
Use PT Series Part Number Crossover Tool to convert from 100, 200, 300 and 615 Series transmitters to new PT Series transmitters.
Type |
N/A Pressure Gauge |
Series |
N/A 800 |
Connection Type |
N/A National Pipe Thread (NPT) |
Cover Ring Material |
N/A 304 Stainless Steel |
Full Scale Accuracy (±) |
N/A 0.25 % |
Connection Size |
N/A 1/4 in |
Lens |
N/A Instrument Glass |
Dial |
N/A Aluminum, white mirrored background with black scale. |
Gauge Size |
N/A 6 in |
Connection Location |
N/A Bottom |
Scale Option |
N/A psi Single Scale |
Case Material |
N/A 304 Stainless Steel |
Measuring Element |
N/A Beryllium copper Bourdon tube to 1,000 psi, 316 SS Bourdon tube 1,500 psi to 6,000 psi |
Connection Material |
N/A Brass |
Case Type |
N/A Stainless Steel (SS) Case, Bottom Connection |
Media Temperature Range1 | N/A -40 to 180 ºF-40 to 80 ºC |
Ambient Temperature Range2 | N/A -40 to 140 ºF-40 to 60 ºC |
Pressure Range |
N/A 0 to 30 psig |
Pointer |
N/A Adjustable Knife-Edge Pointer |
Movement |
N/A Brass with jeweled bearings nickel-silver pinion gear and shafts |
Figure Primary Scale Pressure |
N/A 2 psi |
Graduation Primary Scale Pressure |
N/A 0.1 psi |
Warranty |
N/A 1 yr |
Options |
N/A 304 Stainless Steel (SS) Rear Flange |
Related Items |
N/A
204-MFS, 1/2" National Pipe Thread (NPT), Male x Female, 316 Stainless Steel, 0.187" Orifice Hard Seat Multiport Valve 2070-MFS, 1/2" National Pipe Thread (NPT), Male x Female, 316 Stainless Steel, Hard Seat, 0.187" Orifice Block and Bleed 2-Valve 404-MFS, 1/2" National Pipe Thread (NPT), Male x Female, 316 Stainless Steel, Hard Seat, 0.187" Orifice Needle Valve 704-MFS, 1/2" National Pipe Thread (NPT), Male x Female, 316 Stainless Steel, Soft Seat, 0.187" Orifice Block and Bleed Valve |
Application |
N/A
|
Working Pressure Limitations |
N/A
Static Pressure The working pressure, where no sharp fluctuations occur, should be limited to 100% of the dial range. NOTE: 800 Series Precision Test gauges are not intended for dynamic applications. |
Prop 65 Warnings |
N/A
WARNING: This product can expose you to chemicals including Lead and Nickel, which are known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
|
Caution |
N/A
Operating conditions including, but not limited to, system pressure, media compatibility and ambient conditions must be considered when selecting gauges and accessories. Improper selections and use of gauges could possibly cause gauge failure and lead to possible property damage or personal injury. Refer to the American National Standard ANSI B40.1 for the correct selection and use of gauges. A copy of this standard may be obtained from the American Society of Mechanical Engineers, United Engineering Center, 345 East 47th Street, New York, NY 10017. In keeping with and for purposes of product and/or manufacturing process improvements, NOSHOK, Inc. reserves the right to make design changes without prior notice. |
Limitations Which Apply Are |
N/A
|
Installed Options and Accessories |
N/A
Available
|
Uninstalled Options and Accessories |
N/A
|
Assembly Options |
N/A NOSHOK offers multiple assembly options to meet your specific system requirements. Consult factory to discuss your customized solution. |
Diaphragm Seals |
N/A NOSHOK offers a wide variety of standard & elevated pressure, and reduced pressure replaceable and non-replaceable diaphragm seals, designed to isolate and/or protect the pressure measuring instrument from corrosive, viscous, contaminated or very high temperature process media. |
Distribution Manifolds |
N/A NOSHOK's Distribution Manifolds connect multiple instruments to one diaphragm seal, eliminating multiple connections that lead to possible leak paths. |
Metric Dials and Customized Special Dials |
N/A
|
Pressure Snubbers - Piston Type |
N/A
|
Pressure Snubbers - Sintered Type |
N/A
|
Steam Syphons |
N/A
|