Type |
N/A LevelBAR Electronic Column Indicator |
Series |
N/A 5111 |
Environmental Operating Temperature [Min] |
N/A -1 ºC30 ºF |
Environmental Operating Temperature [Max] |
N/A 49 ºC120 ºF |
Transmitter Full Scale Accuracy |
N/A 0.5 % |
Full Scale Output (FSO) Direct Current |
N/A 4 to 20 mA |
Maximum Load Capacity at 40 V Direct Current Voltage (External Power) |
N/A 1400 O |
Maximum Load Capacity at 24 V Direct Current Voltage (Internal Power) |
N/A 600 O |
Physical Data |
N/A 304 Stainless Steel housing and cover with gasket seal, transparent polycarbonate display window. Intended to meet splash proof requirements. The use of waterproof conduit hubs or fittings may be required during installation. Not rated for continuous explosion hazard areas. |
Frequency [Max] |
N/A 60 Hz |
Frequency [Min] |
N/A 50 Hz |
Display Length |
N/A 10 in254 mm |
Power Output |
N/A 24 V |
Nominal Pressure (Min) |
N/A 0 psid |
Nominal Pressure (Max) |
N/A 15 psid |
Display Accuracy/Resolution |
N/A 1 % |
Alternating Current (AC) Voltage |
N/A 100 to 240 V |
Display Type |
N/A 100-Segment Light Emitting Diode (LED) |
Full Scale Depth (Min) |
N/A 0 to 207 inH2O0 to 5.3 mH2O |
Full Scale Depth (Max) |
N/A 0 to 415 inH2O0 to 10.5 mH2O |
Input |
N/A Pneumatic Input with Internal D/P Transmitter |
Nominal Input Impedance (Resistance) |
N/A 150 O |
Overload Current Limiting |
N/A Automatic trip/reset circuit using PTC (positive temperature coefficient) resistor holds maximum current level within operating range of internal components. |
Pressure Ranges (D/P Option Models) |
N/A 0 - 5, -10, -15, -30, -50 psid fixed range transmitter element (accepts clean, dry air/gas pressure input). Maximum pressure rating is 300% FS. Minimum pressure range for proper operation is 50% of nominal. The lowest acceptable input pressure range is 0 - 2.5 psid using the 0 - 5 psid transducer element. |
Signal Input (Electronic Input Model) |
N/A 4 to 12 mA DC (Minimum) to 4 to 20 mA DC Full Scale Range |
Application |
N/A
|
Replacement Fuses/Terminal Connectors |
N/A
7469-21-0: 500 mA 250 V Fast Acting (AGC) Fuse, 0.25 x 1.25 in, Replacement 5-Pack 5858-2: 2-Terminal Connector 5858-3: 3-Terminal Socket Connector |
Scale Ordering Information |
N/A
A standard 1-unit scale is included in the package price listed for LevelBAR models. The LevelBAR scale is custom marked in any specified unit of measurement (gallons, liters, barrels, etc.). An optional 2-unit scale can combine any two different measurements such as weight and volume, or volume and depth. When ordering, the following information must be supplied. If replacing an existing scale, also provide the original serial number.
Tank Calibrations - The LevelBAR indicator should not be used for physical water calibrations of tanks. Use of a pressure standard or other high accuracy pressure measurement instrument is recommended. |
Replacement Scales |
N/A
5980-11: Replacement 1-Unit LevelBAR Scale Assembly 5980-21: Replacement 2-Unit LevelBAR Scale Assembly |
LevelBAR Custom Scale |
N/A Scales are calculated for the individual tank geometry and product specific gravity permitting the LevelBAR system to be used for almost any kind and shape of tank, including horizontal cylindrical storage tanks. Custom marked in any specified unit of measurement (gallons, liters, pounds, etc.), it may also be ordered with an optional 2-unit scale that combines any two different measurements of weight, volume, or depth. Factory calibration allows full scale ranging of the display when maximum input is less than 20 mA to eliminate field span adjustment of the level transmitter. |
LevelBAR Indicator |
N/A
These versions incorporate the D/P option with an internal pressure transducer. Select the pressure range based on the depth and density of liquid contents. A factory calibration scaling adjustment provides full scale display for most applications. These models include 1/4" tube fitting, watertight cord grip (strain relief) connectors, factory calibration, and custom marked scale. Pneumatic - Pressure signal from Bubbler or Air-Driven Diaphragm Sensor is transmitted to LevelBAR (D/P Option). |
LevelBAR Direct Pressure Input (D/P Option) |
N/A
The available LevelBAR D/P option incorporates a precision pressure transducer for pneumatic-to-current (P/I) conversion, allowing the unit to read pressure directly. This available feature permits bubbler systems to be retrofitted with an electronic LevelBAR indicator, eliminating older manometer type gages. Additionally, the indicator generates a compatible 4-20 mAdc output signal (In some instances, the maximum applied pressure may not provide full scale output and result in less than 20 mAdc.) that may be linked to a PLC or other control/monitoring system. LevelBAR indicators are suitable for use with air operated level sensors (force balance diaphragm or air purge bubbler). These systems enable the pneumatic output signal to be routed directly into the indicator and displayed on the LED column. The LevelBAR can also be used to retrofit systems that included mercury gauges or other fluid filled tank level gauges. It offers a standardized mounting footprint when installing the indicators even when depths vary from tank to tank. Other advantages include reduced maintenance and eliminating concerns about reliability.
|
Additional Information |
N/A
The multi-segment LED column provides graphic indication of tank levels for operations personnel. Rugged stainless steel housing resists corrosion affording protection for the internal electronics and is intended to meet splash proof requirements. A backlighting feature makes the scale readable in the absence of ambient illumination while the red LED column is viewable even in bright daylight. LevelBAR indicators can be used singly or in multiples to provide additional display locations simply by extending the signal loop circuit. This tank level display takes the guesswork out of installation and troubleshooting with practical elements such as loop circuit status (on/off/open) and fault (short) indicators. With the available differential pressure (D/P) option, the LevelBAR indicator can be used as a direct replacement for mercury tank gauges or other fluid-filled manometers. |
Product Warranty |
N/A
All KING-GAGE®, a NOSHOK Company products are guaranteed to be free from defects in material and workmanship for one (1) year from the date of purchase. Repair Services Warranty: 90 days limited warranty. |
Prop 65 Warnings |
N/A
WARNING: This product can expose you to chemicals including Lead and Nickel, which are known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
|