• High visibility tank level display
  • Individually calibrated indicator scale
  • Electronic (4 mA to 20 mA) or pneumatic input
  • Direct replacement for manometer gauges
Unit of Measure

Specifications

Type

N/A LevelBAR Electronic Column Indicator

Series

N/A 5111

Environmental Operating Temperature [Min]

N/A -1 ºC30 ºF

Environmental Operating Temperature [Max]

N/A 49 ºC120 ºF

Transmitter Full Scale Accuracy

N/A 0.5 %

Full Scale Output (FSO) Direct Current

N/A 4 to 20 mA

Maximum Load Capacity at 40 V Direct Current Voltage (External Power)

N/A 1400 O

Maximum Load Capacity at 24 V Direct Current Voltage (Internal Power)

N/A 600 O

Physical Data

N/A 304 Stainless Steel housing and cover with gasket seal, transparent polycarbonate display window. Intended to meet splash proof requirements. The use of waterproof conduit hubs or fittings may be required during installation. Not rated for continuous explosion hazard areas.

Frequency [Max]

N/A 60 Hz

Frequency [Min]

N/A 50 Hz

Display Length

N/A 10 in254 mm

Power Output

N/A 24 V

Nominal Pressure (Min)

N/A 0 psid

Nominal Pressure (Max)

N/A 15 psid

Display Accuracy/Resolution

N/A 1 %

Alternating Current (AC) Voltage

N/A 100 to 240 V

Display Type

N/A 100-Segment Light Emitting Diode (LED)

Full Scale Depth (Min)

N/A 0 to 207 inH2O0 to 5.3 mH2O

Full Scale Depth (Max)

N/A 0 to 415 inH2O0 to 10.5 mH2O

Input

N/A Pneumatic Input with Internal D/P Transmitter

Nominal Input Impedance (Resistance)

N/A 150 O

Overload Current Limiting

N/A Automatic trip/reset circuit using PTC (positive temperature coefficient) resistor holds maximum current level within operating range of internal components.

Pressure Ranges (D/P Option Models)

N/A 0 - 5, -10, -15, -30, -50 psid fixed range trans­mitter element (accepts clean, dry air/gas pressure input). Maximum pressure rating is 300% FS. Minimum pressure range for proper opera­tion is 50% of nominal. The lowest acceptable input pressure range is 0 - 2.5 psid using the 0 - 5 psid transducer element.

Signal Input (Electronic Input Model)

N/A 4 to 12 mA DC (Minimum) to 4 to 20 mA DC Full Scale Range

Application

Application

N/A
  • Tank level Indication
  • Marine
    • Tank Level, Draft Monitoring and Vessel Management
    • Ballast Water
    • Fuel Oil
    • Fresh Water
    • Lube Oil
    • Brine
    • Preload
    • Holding

Options and Accessories

Replacement Fuses/Terminal Connectors

N/A 7469-21-0: 500 mA 250 V Fast Acting (AGC) Fuse, 0.25 x 1.25 in, Replacement 5-Pack

5858-2: 2-Terminal Connector

5858-3: 3-Terminal Socket Connector

Scale Ordering Information

N/A A standard 1-unit scale is included in the package price listed for LevelBAR models. The LevelBAR scale is custom marked in any specified unit of measurement (gallons, liters, barrels, etc.). An optional 2-unit scale can combine any two different measurements such as weight and volume, or volume and depth.

When ordering, the following information must be supplied. If replacing an existing scale, also provide the original serial number.
  • Specific Gravity (Density) of Tank Contents
  • Product Designation (tank contents)
  • Description of Tank (e.g., Walker 6000 gal silo)
  • Tank Dimensions
  • Tank Name (to be noted on scale)
  • Pressure Range of transmitter or LevelBAR
  • Scaling Units (e.g., lbs, gals, liters, etc.)
Scales are calculated for the individual tank geometry and product specific gravity. Scaling adjustments setting (REF HIGH) is noted on the side of the scale. This is used to set the full scale display based on pressure range of the transmitter and maximum pressure created by depth of tank product.

Tank Calibrations - The LevelBAR indicator should not be used for physical water calibrations of tanks. Use of a pressure standard or other high accuracy pressure measurement instrument is recommended.

Replacement Scales

N/A 5980-11: Replacement 1-Unit LevelBAR Scale Assembly

5980-21: Replacement 2-Unit LevelBAR Scale Assembly

Features

LevelBAR Custom Scale

N/A Scales are calculated for the individual tank geometry and product specific gravity permitting the LevelBAR system to be used for almost any kind and shape of tank, including horizontal cylindrical storage tanks. Custom marked in any specified unit of measurement (gallons, liters, pounds, etc.), it may also be ordered with an optional 2-unit scale that combines any two different measurements of weight, volume, or depth. Factory calibration allows full scale ranging of the display when maximum input is less than 20 mA to eliminate field span adjustment of the level transmitter.

LevelBAR Indicator

N/A These versions incorporate the D/P option with an internal pressure transducer. Select the pressure range based on the depth and density of liquid contents. A factory calibration scaling adjustment provides full scale display for most applications. These models include 1/4" tube fitting, watertight cord grip (strain relief) connectors, factory calibration, and custom marked scale.

Pneumatic - Pressure signal from Bubbler or Air-Driven Diaphragm Sensor is transmitted to LevelBAR (D/P Option).

LevelBAR Direct Pressure Input (D/P Option)

N/A The available LevelBAR D/P option incorporates a precision pressure transducer for pneumatic-to-current (P/I) conversion, allowing the unit to read pressure directly. This available feature permits bubbler systems to be retrofitted with an electronic LevelBAR indicator, eliminating older manometer type gages. Additionally, the indicator generates a compatible 4-20 mAdc output signal (In some instances, the maximum applied pressure may not provide full scale output and result in less than 20 mAdc.) that may be linked to a PLC or other control/monitoring system.

LevelBAR indicators are suitable for use with air operated level sensors (force balance diaphragm or air purge bubbler). These systems enable the pneumatic output signal to be routed directly into the indicator and displayed on the LED column. The LevelBAR can also be used to retrofit systems that included mercury gauges or other fluid filled tank level gauges. It offers a standardized mounting footprint when installing the indicators even when depths vary from tank to tank. Other advantages include reduced maintenance and eliminating concerns about reliability.
  • Simple graphic display of level
  • LED column 10 in./254 mm
  • Backlit scale for nighttime viewing
  • Replaces mercury gauges

Additional Information

Additional Information

N/A
  • High Visibility Tank Level Display
  • Individually Calibrated Indicator Scale
  • Electronic (4-20 mA) or Pneumatic Input
  • Direct Replacement for Manometer Gauges
The KING-GAGE® LevelBAR (v2) delivers a bright multi-segment LED column to graphically indicate tank level. The system monitors tank level by reading the proportional 4-20 mA signal output from compatible two wire level transmitters. The available D/P option alternatively accommodates direct pneumatic pressure measurement.

The multi-segment LED column provides graphic indication of tank levels for operations personnel. Rugged stainless steel housing resists corrosion affording protection for the internal electronics and is intended to meet splash proof requirements. A backlighting feature makes the scale readable in the absence of ambient illumination while the red LED column is viewable even in bright daylight.

LevelBAR indicators can be used singly or in multiples to provide additional display locations simply by extending the signal loop circuit. This tank level display takes the guesswork out of installation and troubleshooting with practical elements such as loop circuit status (on/off/open) and fault (short) indicators. With the available differential pressure (D/P) option, the LevelBAR indicator can be used as a direct replacement for mercury tank gauges or other fluid-filled manometers.

Warnings

Product Warranty

N/A All KING-GAGE®, a NOSHOK Company products are guaranteed to be free from defects in material and workmanship for one (1) year from the date of purchase.

Repair Services Warranty: 90 days limited warranty.

Prop 65 Warnings

N/A
WARNING: This product can expose you to chemicals including Lead and Nickel, which are known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov

Note

Note

N/A Scale information must be supplied when order is placed.