• Flush mounting sanitary design
  • Pneumatic output, or electronic (4 mA to 20 mA) output when paired with KING-GAGE® 860 Series Sensor Control or 868 Series D/P Purge Control
  • Field replaceable sensor diaphragm
  • Stable output during CIP cycles
  • FDA-approved silicone rubber O-ring seal forms a crevice-free, positive compression seal between the diaphragm probe and mounting shell
  • Flush weld mountings are available with product contact surfaces of
Unit of Measure

Specifications

Type

N/A AcraSensor Level Transmitter

Series

N/A 450

Mounting Application

N/A Short mount (single wall, non-insulated tank) Use with weld-in shell 1777-2 or to fit 2" sanitary clamp fitting adapter

Output/Connections

N/A Electronic (4-20 mA), 1/4 in. Tube Swagelok

Wetted Material

N/A Silicone rubber O-ring (certified to 3-A 18-03) Titanium diaphragm, commercially pure (certified to 3-A 74-06).
Optional diaphragm materials: Silicone rubber
Fittings/adapter wetted surfaces are typically 316L Stainless Steel (as supplied by KING-GAGE®).

Air Consumption

N/A Total air consumption of sensor package (including sensor control) is less than 10 ft³/h or 0.16 scfm.

Voltage Supply Requirements

N/A 14 Vdc to 40 Vdc (Unregulated) to Power Signal Loop Using 868 D/P Sensor Control.

Pressure Range

N/A Pneumatic, Includes SafeGard Option (25 to 650 in. Range)

Note for Pressure Range

N/A Sensor is a 1:1 pneumatic force balance diaphragm capable of repeating hydrostatic pressure as an equivalent pneumatic output. Functional pressure limit is 130 psi (or 3,598 inches at 1.00 specific gravity) determined by supply pressure (35 to 150 psi) less 20 psi.

Sensitivity

N/A Less than 0.001 psi (± 0.02 inH2O/ 0.508 mmH2O)

Repeatability

N/A Better than ± 0.002 psi (± 0.05 inH2O /1.27 mmH2O)

Note for Span Adjustment Range

N/A IMPORTANT! Accuracy, linearity, and non-repeatability values are based on nominal range.

Operating Limit (Maximum Pressure)

N/A Pressure above 300% nominal range (over range) will result in damage to the transmitter (200% may cause a shift in calibration). Burst pressure is 200 psi and will cause catastrophic and physical failure of the pressure element.

Accuracy

N/A ± 0.007 psi (± 0.2 inH2O/0.508 mmH2O)

Depth Measurement [Max]

N/A 1,937 inH2O49.0 mH2O

Transmitter Signal Loop

N/A As is typical of two wire transmitters, electrical power to the signal loop is supplied through twisted pair cable. Excitation voltage (+V) is provided by the receiving device or from a separate Vdc power supply installed in the loop. The actual voltage necessary for operation of the system will depend upon the overall load resistance of the loop. In most applications, the KING-GAGE® Digital Indicator or LevelBAR indicator provide 24 Vdc excitation to power the transmitter signal loop.

Temperature Ranges

Operating Temperature [Min]

N/A 0 ºC30 ºF

Operating Temperature [Max]

N/A 149 ºC300 ºF

Pressure Ranges

Pneumatic Operation

N/A AcraSensor diaphragm unit is paired with a Sensor Control regulator to provide the continuous 1 CFH (cubic foot per hour) air flow to the underside of the diaphragm to attain pneumatic balance pressure. Constant back pressure regulation maintains a linear 1:1 response throughout the entire operating range to ensure pneumatic pressure is directly equivalent to the hydrostatic head of liquid in the tank.

The AcraSensor + 860 Sensor Control combination affords rugged durability and simple maintenance. Used in conjunction with compatible indicators (such as the KING-GAGE® LevelBAR), this liquid level sensor can provide reliable operation in even the most challenging environments. Alternative configurations include substitution of a molded silicone rubber diaphragm when appropriate.

Compressed Air Supply

N/A Operation of the AcraSensor requires a compressed air/gas source providing supply pressure of 35-150 psig (2.4-10.3 bar). Pneumatic components require instrumentation-grade compressed air that is clean, dry and oil-free to ensure trouble free performance.

Application

Application

N/A
  • Food and beverage
  • Chemical processing
  • Power generation
  • Pharmaceutical
  • Water and wastewater
  • Pulp and paper
  • Marine

Tank Gauging Applications

N/A Acting on the force balance principle, a pneumatic sensor generates air pressure equivalent to the hydrostatic force created by liquid depth. This pneumatic pressure is directed into the D/P transmitter. Pressure is converted to an electronic 4-20 mAdc output proportional to liquid depth. Twisted pair cabling is used to form the signal loop through which the remote indicator receives the transmitter output. The KING-GAGE® indicators shown in these system views correlate the transmitter output to total volume or weight of tank contents. This is determined by using a pressure versus capacity listing corresponding to the actual tank dimensions.

The standard AcraSensor + 868 D/P Sensor Control combination combines accuracy, rugged durability, reduced maintenance and long term stability for liquid level tank gauging. Its integral transmitter converts the pneumatic isolation pressure to a proportional 4-20 mAdc signal suitable for process control or remote level indication.

Options and Accessories

Service and Replacement Parts

N/A 7333-10: Replacement filter element
7334-2: SafeGard service kit
7334-9: Air control section kit
7879-20: Vent generator
9316-4: O-ring for D/P top plate (868)

Flush Weld Shells and Retrofit Adapters

N/A Polished stainless steel tank mounting is offered in two lengths for insulated wall vessels and a short version for single wall (non-insulated) tanks. We offer adapter inserts that permit retrofit installation of the AcraSensor unit within an existing sanitary mounting from other manufacturers (e.g., Tank Mate, Cherry Burrell, Anderson Instrument). There is also an adapter for 2" Tri-Clamp quick release fittings available for direct mounting of the AcraSensor unit.

860 Series Sensor Control

N/A
  • Better than ±0.02% span repeatability
  • Also offered in hybrid versions combined with an air flow control for pneumatic sensors or downpipe (bubbler) systems

868 Series Differential Pressure Sensor Control

N/A
  • 4 mA to 20 mA output for broad compatibility
  • Better than ±0.02% span repeatability
  • Piezoresistive technology for enhanced stability
  • Also offered in hybrid versions combined with an air flow control for pneumatic sensors or downpipe (bubbler) systems

3-Tube Cable (Replacement or Non-Standard Lengths)

N/A Use for interconnection between AcraSensor and Sensor Control

1766-15: Cable, 15 ft (4.5 m) total length, trimmed
8014-X: Cable quantity discount, minimum purchase 50 feet, X = Total Length

Acrasensor Replacement/Service Parts

N/A 1751: Clamp Nut, 304 SS
1752-22-3: Acrasensor Connector Assembly (old)
1752-30-3: Acrasensor Connector Assembly (new)
6554-40: O-Ring, Silicone Rubber (Standard)

1747-40: O-Ring, Fluorocarbon (FKM) (Special Application) Consult Factory
1786-40: O-ring, PTFE (Special Application) Consult Factory
6554-44: O-Ring, Oversized (Special Application) Consult Factory

5611-21: Acrasensor Housing (Short)
5611-22: Acrasensor Housing (Standard)
5611-23: Acrasensor Housing (Long)
7859-1: Post Assembly Short
7859-2: Post Assembly Standard
7859-3: Post Assembly Long
7899-3: Body (For Rubber Diaphragm)

AcraSensor Diaphragms

N/A 7904-1: AS Titanium Diaphragm Assembly with Silicone O-Ring
S-1618-3: AS Titanium Diaphragm Assembly with PTFE O-Ring
6448-12-5: Silicone (White Rubber) Diaphragm 2-Pack (Rubber Diaphragm is Used with 7899-3 Body)

Diaphragm Sensing Units

N/A Acrasensor (Titanium with Silicone Rubber O-Ring)
8999-1-4: Short Length, Titanium Diaphragm
8999-2-4: Standard Length, Titanium Diaphragm
8999-3-4: Long Length, Titanium Diaphragm

Acrasensor (Titanium Diaphragm with PTFE O-Ring)
8999-1-5: Short Length, S-1618-3 Diaphragm
8999-2-5: Standard Length, S-1618-3 Diaphragm
8999-3-5: Long Length, S-1618-3 Diaphragm

Standard Length - Use with 1777 weld-in shell for single wall or insulated wall (up to 4.5" thick; also fits adaptors for non-KING-GAGE® "short" shells.

Long Length - Use with1777-3 weld-in shell for Crepaco silo or other thicker wall (up to 6.75") vessels; also fits adapters for non-KING-GAGE® standard (insulated wall) shells.

Short Length - For single wall (non-insulated) tanks; fits KING-GAGE® short shell.

Note: Laser-fused Titanium Diaphragm units are generally recommended for both sanitary and industrial applications.

Differential Pressure Capsule Replacement

N/A Replacement D/P capsule (sensor/transmitter assembly) for Model 868 D/P Sensor Controls. Signal input connection is 1/4 NPT - part numbers below do not include fitting.

S-1565-1-5-1: D/P Replacement Capsule Kit (0-2 psid)
S-1565-1-0-1: D/P Replacement Capsule Kit (0-5 psid)
S-1565-1-1-1: D/P Replacement Capsule Kit (0-10 psid)
S-1565-1-2-1: D/P Replacement Capsule Kit (0-15 psid)
S-1565-1-3-1: D/P Replacement Capsule Kit (0-30 psid)
S-1565-1-4-1: D/P Replacement Capsule Kit (0-50 psid)

Features

Reliable Liquid Level Measurement

N/A These unique pneumatic sensors detect level by sensing hydrostatic pressure created by liquid depth. Acting on the force balance principle, a sensitive diaphragm is exposed to the liquid contents of the tank. Compressed air within the sensor creates a pneumatic pressure that balances the force of the liquid acting against the diaphragm. This compensating pneumatic pressure is directly proportional to the depth of liquid in the tank.

The pneumatic force balance technique is virtually free from long term drift, hysteresis and temperature sensitivity unlike strain measurements (including load cells) since mechanical deflection of the diaphragm is not directly measured. The compensating pneumatic balance pressure which effectively isolates the sensor is converted to a proportional 4-20 mAdc signal or can be directly output to a suitable gauge. In actuality, the sensor is measuring the mass (weight) of fluid matter that can be correlated to units of standardized volume.

Sanitary Configurations

N/A These sensors are especially suited to a wide range of sanitary applications on storage or continuous processing vessels. The unique AcraSensor titanium diaphragm satisfies the 3-A Sanitary Standard and is appropriate for flush mount, clean-in-place installation. An FDA-approved silicone rubber O-ring seal forms a crevice-free, positive compression seal between the diaphragm probe and mounting shell. Flush weld mountings are available with product contact surfaces of 316L stainless.

The level sensor can be alternatively supplied with silicone rubber diaphragm surface. This molded elastomer provides additional material compatibility options. (Quick disconnect sanitary adapters are also available for sensor mounting.)

Conformance to Sanitary Standard

Conformance to Sanitary Standard

N/A The AcraSensor diaphragm is authorized to carry the 3-A symbol and conforms to 3-A Sanitary Standards 74-06, Sensors and Sensor Fittings and Connections.

Installation Options

Installation Options

N/A For sanitary flush mount installation, shell is installed through the side of the tank and flush-welded to inner liner. AcraSensor uses a one-piece titanium diaphragm mounted onto a post assembly inserted into the tank shell and clamped into place.

Additional Information

Additional Information

N/A Meets current standards for 3A
  • Flush mounting sanitary design
  • Electronic (4-20 mA) output or pneumatic output
  • Field replaceable sensor diaphragm
  • Stable output during
  • CIP cycles

Warnings

Product Warranty

N/A All KING-GAGE®, a NOSHOK Company products are guaranteed to be free from defects in material and workmanship for one (1) year from the date of purchase.

Repair Services Warranty: 90 days limited warranty.

Prop 65 Warnings

N/A
WARNING: This product can expose you to chemicals including Lead and Nickel, which are known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov

Note

Note

N/A AcraSensor packages listed include diaphragm sensing unit, 860 Series Sensor Control or 868 Series D/P Purge Control, Swagelok or PolyFlo fittings, and bundled 3-tube airline cable.

Application Notes

N/A Location of the Sensor Control air flow regulator should be generally no greater than 50 feet (15 m) from the AcraSensor unit. The level sensor includes 15 feet (4.5 m) of bundled 3-tube cable for the pneumatic interconnections with the Sensor Control. Distances greater than 50 feet may increase response time of the sensor to changes in liquid level in the tank.