Type |
N/A LevelBAR Electronic Column Indicator |
Series |
N/A 5111 |
Environmental Operating Temperature [Min] |
N/A -1 ºC30 ºF |
Environmental Operating Temperature [Max] |
N/A 49 ºC120 ºF |
Full Scale Output (FSO) Direct Current |
N/A 4 to 20 mA |
Physical Data |
N/A 304 Stainless Steel housing and cover with gasket seal, transparent polycarbonate display window. Intended to meet splash proof requirements. The use of waterproof conduit hubs or fittings may be required during installation. Not rated for continuous explosion hazard areas. |
Frequency [Max] |
N/A 60 Hz |
Frequency [Min] |
N/A 50 Hz |
Display Length |
N/A 10 in254 mm |
Power Output |
N/A 24 V |
Nominal Pressure (Min) |
N/A 0 psid |
Nominal Pressure (Max) |
N/A 5 psid |
Display Accuracy/Resolution |
N/A 1 % |
Alternating Current (AC) Voltage |
N/A 100 to 240 V |
Display Type |
N/A 100-Segment Light Emitting Diode (LED) |
Full Scale Depth (Min) |
N/A 0 to 69 inH2O0 to 1.8 mH2O |
Full Scale Depth (Max) |
N/A 0 to 138 inH2O0 to 3.5 mH2O |
Note for Depth |
N/A Typical applications should not exceed 10 feet/3 meters. Greater depth will require excessive hand pumping. |
Input |
N/A Hand Pump Operation for Downpipe Purge |
Nominal Input Impedance (Resistance) |
N/A 150 O |
Overload Current Limiting |
N/A Automatic trip/reset circuit using PTC (positive temperature coefficient) resistor holds maximum current level within operating range of internal components. |
Signal Input (Electronic Input Model) |
N/A 4 to 12 mA DC (Minimum) to 4 to 20 mA DC Full Scale Range |
Application |
N/A
|
Replacement Fuses/Terminal Connectors |
N/A
7469-21-0: 500 mA 250 V Fast Acting (AGC) Fuse, 0.25 x 1.25 in, Replacement 5-Pack 5858-2: 2-Terminal Connector 5858-3: 3-Terminal Socket Connector |
Scale Ordering Information |
N/A
A standard 1-unit scale is included in the package price listed for LevelBAR models. The LevelBAR scale is custom marked in any specified unit of measurement (gallons, liters, barrels, etc.). An optional 2-unit scale can combine any two different measurements such as weight and volume, or volume and depth. When ordering, the following information must be supplied. If replacing an existing scale, also provide the original serial number.
Tank Calibrations - The LevelBAR indicator should not be used for physical water calibrations of tanks. Use of a pressure standard or other high accuracy pressure measurement instrument is recommended. |
Replacement Scales |
N/A
5980-11: Replacement 1-Unit LevelBAR Scale Assembly 5980-21: Replacement 2-Unit LevelBAR Scale Assembly |
LevelBAR Custom Scale |
N/A Scales are calculated for the individual tank geometry and product specific gravity permitting the LevelBAR HP system to be used for almost any kind and shape of tank, including horizontal cylindrical storage tanks. Each is custom marked in any specified unit of measurement (gallons, pounds, liters, short tons, etc.). An optional 2-unit scale can be specified to combine any two different measurements such as weight and volume, or volume and depth. |
Additional Information |
N/A
All LevelBAR indicators include a custom graduated tank level scale in units of total volume, weight or depth of the liquid contents. A backlighting feature makes the scale readable in the absence of ambient illumination while the red LED column is viewable even in bright daylight. LevelBAR HP indicator can be used as a direct replacement for mercury tank gauges or other fluid-filled manometers. This unit is well adapted for isolated tank level gauging applications where continuous readings are not required. Rugged stainless steel housing resists corrosion while affording suitable protection for the internal electronics and intended to meet splash proof requirements. The use of waterproof conduit hubs or fittings may be required during installation. Not rated for continuous explosion hazard areas. (Requires 100-240 Vdc 50/60 Hz power.) Simple proven principle employs a hand pump to supply air which fills the downpipe inserted into the tank liquid. The resulting pressure created (within the pipe) and sensed by the indicator is directly proportional to liquid depth. |
Product Warranty |
N/A
All KING-GAGE®, a NOSHOK Company products are guaranteed to be free from defects in material and workmanship for one (1) year from the date of purchase. Repair Services Warranty: 90 days limited warranty. |
Prop 65 Warnings |
N/A
WARNING: This product can expose you to chemicals including Lead and Nickel, which are known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
|